High-temperature furnace
HTO-34

High-temperature furnace HTO-34
for material testing in air and a heating zone length of 30 cm
General Features
The high-temperature furnace HTO-34 is characterized by an excellent temperature distribution, which is achieved by its 3-zone version.Each zone can be controlled independently, allowing precise temperature control, especially when using an appropriate temperature controller.
The furnace is made more energy-efficient through the use of a lightweight, thermally insulating and asbestos-free aluminum silicate wool.
The housing of the furnace is made of stainless steel, which increases the longevity of the appliance.
In addition, the furnaces are already equipped to accommodate a length change sensor(HT extensometers) and diametrical HT extensometers through the cross slots at the front and rear. This makes it easy to integrate measuring devices for monitoring material changes.
| Technical information | ||
|---|---|---|
| Mass and weight | ||
| Heated zone (length) | 3 zone: Length per zone: | Per zone approx. 100 mm (vertical) Can be controlled independently |
| Furnace external dimensions | Ø: Height: | Approx. 295 mm Approx. 461 mm (405 mm – furnace dimensions without furnace rack) |
| Heating room internal dimensions | Ø: Height: | Approx. 100 mm Approx. 300 mm |
| Extensometer openings | Height x Width: | cross-slots fornt rear: 60 x 25 mm (horizontal) 25 x 100 mm (longitudinal) |
| Feed-through opening | Ø top and bottom: | Approx. 50 mm |
| Weight | Approx. 45 kg | |
| Thermal values | ||
| Temperature range | (Indoor) room temperature up to 1150°C | |
| Permissible heating element surface temperature | Max. 1250 ° C | |
| Heating rate | max. 20 K/min. | |
| Materials | ||
| Heating conductor | • Fibrothal modules with A1 resistance coils • Horizontally positioned and free-radiating | |
| Thermocouple | • NiCR-Ni, Type N | |
| Thermal insulation | • Polycrystalline AI2O3- Wool | |
| Housing | • Stainless steel • Double-walled design • Openable • Reduction of the outer wall temperature by convection air cooling between the insulation and the housing wall • Perforated sheet outer casing for additional heat dissipation • Housing can be opened with a side opening in the separation area for the passage of the HT extensometer or a sample thermocouple unit (on the back) • Device underneath the furnace for locking one of the two furnaces halves | |
| Electrics | ||
| Power/Connection | Approx. 0,6 KW per zone / 3 x 34 V, 50 Hz Safety switch for OFF when opening the furnace |
Features
- Double-walled construction – convection cooling reduces the outside wall temperature.
- The furnace stands on furnace supports so that the furnace can be fully opened.
- The MSRA regulator uses transformers to provide the required high current and reduce the voltage. Thanks to the lower voltage, the oven's lifespan is extended.
- Folding furnace. Fast and easy sample loading of samples possible.
- 2 double Cross-slots (front and rear for additional diametral extensometers)
- Durable embedded horizontal heating elements.
- Furnace bore optimized for a wide range of adapters and test fixtures without causing large heat losses.
- Lightweight and asbestos-free insulation material to minimize power consumption.
- Stainless steel housing.
- All electrical connections are made to the MSRA controller.
- Front cutout allows access to high-temperature extensometers.
- Uniform temperature gradient through 3 independent heating zones.
- Five feed-throughs for thermocouples, one on the right and one on the left, three of which are in the middle heating zone.
- Movable thermocouple arrays for temperature are available as an option.
- Adjustable stainless steel latches for easy locking of the furnace holder during the test.
- Upper and lower end plates with tie rod cutouts are manufactured to fit the test device.
- OFF safety switch when opening the furnace.
- 2 cross-slot extensometer openings:
For length change extensometers and for diametrical extensometers.
Application
- Heating zone length of 300 mm.
- Suitable for high-temperature tensile tests, compression tests, cyclic fatigue tests and fracture mechanics tests.
- For tensile specimens B from M6 to M20 (according to DIN 50125).
- Load rods for tensile specimens up to M20 available.
GET IN CONTACT
WITH US!